Model : INCO 2/108
- Volume : 108 liters - Temperature : Operating range from 7oC above ambient up to 45o C Temperature fluctuations with time (to DIN 58945) £ ±0.10C Temperature variation in chamber (to DIN 58945) £± 0.50C Electronic microprocessor temperature controller with Pt 100 and auto-diagnostic system - Ventilation : Uniform atmosphere and temperature distribution through enclosed non-turbulent ventilation system, fully covered by the sterilisation process. - Humidity : Adjustable humidity limit control (88-97%) incl.digital indication and auto-diagnostic system with visual and acoustic fault indication (air supply via sterilefilter) ensures rapid reaching of set humidity and short recovery times while avoiding condensate formation. Standard water dish number : 01 - CO2 : Digital electronic CO2 control with autozero, NDIR system and integrated gas cylinder change-over, with autodiagnostic system and acoustic fault indication. Setting accuracy : ±0.1% Adjustment range : 0 up to 20% - Interface : Parallel printer interface for graphic or numerical documenting of temperature, CO2, and relative humidity for EPSON ESC/P2 compatible printers Computer interface RS232 with software “Celsius 2000”for temperature, CO2, and relative humidity. Stainless steel interior (W x D x H) : 560 x 400 x 480 mm Textured stainless steel exterior (rear zinc-plated steel) (W x D x H) : 710 x 590 x 770 mm Fully insulated, heated stainless steel door Extra internal glass door Provision for sliding stainless steel shelves : 04. Standard shelves : 03 Electrical supply 230 V, 50Hz
Other model : 153 liters (INCO 2/153), 246 liters (INCO 2/246) |