The U-BTC is designed for thermal battery testing from individual cells to complete battery systems. It supports a wide range of applications in development, validation and
quality assurance, including test tasks related to standards such as IEC 62660, IEC 62619 and ISO 12405-4.
Typical users include automotive OEMs and suppliers, battery cell manufacturers, energy storage providers, test laboratories and research institutions.
Relevant standards
01 EN 62133 – Safety of secondary cells
02 IEC 60086 – Lithium batteries
03 UL 2054 – Household batteries
04 UL 2580 – Batteries for electric vehicles
05 EN 50604 – Batteries for light electric vehicles
06 ISO 12405 – Testing of battery packs
07 IEC 62660 – Secondary lithium-ion cells for electric vehicles
The Memmert U-BTC is designed for applications up to Hazard Level 4
+ Gas release, mass loss > 50%
+ Electrolyte loss, cell venting, no flame or fire.
Specifications:
– Volume: 108 liters
– Interior dimensions: width 560 x height 480 x depth 400 mm
– Exterior dimensions: width 745 x height 864 x depth 584 mm
– Setting temperature range: 20 – 120ºC
– Working temperature range: at least 10 above ambient temperature to +120°C
– Setting accuracy temperature: up to 99.9 °C: 0.1 / from 100 °C: 0.5
– Max. pressure surge load on the test chamber: 10 mbar
– Actuation pressure of the pressure relief flaps: 5 mbar
– Test specimen type: 18650 cell, 2.2 Ah
– Test specimen: 3
– Max. total gas volume of the test specimen: 33 liters
– Max. air flow through the pressure relief flap: 1200 l/min
– Control unit: TwinDISPLAY. Intuitive display (colour graphic display) with touchscreen.
– Adjustable parameters: temperature
– Adjustable device functions: Flap position, time zones, alarm definition, real-time display of process variables, Network settings (IP, subnet, DHCP), timer function, and much more
– Language setting: German, English, Spanish, French, Polish, Czech, Hungarian, Italian, Chinese
– Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
– SetpointWAIT: process time does not start until the set temperature is reached
– Calibrate: three freely selectable temperature values
– Interfaces: Ethernet LAN, USB
– Door: Fully insulated stainless steel door with a 2-point locking system (compression door lock) and additional tension locks to secure the door
– Convection: natural ventilation
– Fresh air: Admixture of pre-heated fresh air by electronically adjustable air flap
– Exhaust air: vent connection with restrictor flap, electronically adjustable
– Logging: programme stored in case of power failure
– Parameter monitoring: 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error.
– Temperature monitoring: Mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 200C above nominal temperature
– Temperature monitoring: overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display
– AutoSAFETY: additionally integrated over- and undertemperature monitor “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature
– Autodiagnostic system: for fault analysis
– Alarm: visual and acoustic
– Pressure relief: To relieve pressure, the test chamber is fitted with a 120 mm diameter entry port. If excess pressure builds up inside the chamber – for example, due to gas being released from a test battery – the flap opens and relieves the pressure in a controlled manner. It then closes automatically under spring tension.
– Stainless steel interior
– Textured stainless steel casing
– Maximum of grid: 5
– Maximum loading of chamber: 175 kg
– Maximum loading per grid: 20 kg
– Voltage: 230 V, 50/60 Hz, 2800 W
– Ambient temperature: 5 – 400C
– Humidity: max. 80% rh, non-condensing
Included:
+ Works calibration certificate 1000C (Memmert)
+ Stainless steel grid, 2 pcs
+ Manual