Defined air conditions.
Measured. Documented. Traceable.
Designed for the defined air conditions required by ISO 188:2023 Method D and DIN 53508:2024-03 Method D – with device-specific measurement of the air change rate and documented evidence of the airflow conditions.
Precise and verifiable ageing tests
The Memmert UFACR-D is designed for demanding thermal ageing tests in which defined air conditions are just as critical as temperature and test duration. ISO 188:2023 Method D and DIN 53508:2024-03 Method D define specific requirements for air change rate and air speed.
A purpose-engineered interior configuration guides the airflow through the chamber in a controlled manner and creates the defined air conditions required for reproducible testing.
A key advantage for testing laboratories and quality assurance is the device-specific verification of the air change rate. Before delivery, every UFACR-D is measured in the Memmert test field and the result is documented in a device-specific test certificate for the air change rate. This provides clear, traceable documentation for quality management, audits and technical traceability.
Key performance features
- Developed for Method D: ISO 188:2023 Method D and DIN 53508:2024-03 Method D as clear application references.
- Purpose-engineered interior: Guides the airflow through the chamber to create the defined air conditions required for Method D.
- Defined air change rate: ISO 188:2023 Method D requires more than 30 air changes per hour; DIN 53508:2024-03 Method D requires at least 30 air changes per hour. The chamber design is engineered accordingly
- Defined air speed: Method D specifies an air speed of 0.25 to 3.0 m/s at the test pieces
- Documented evidence: Standard documentation includes a device-specific test certificate for the air change rate and a separate air flow certificate.
Specifications:
– Volume: 29 liters
– Interior dimensions: width 314 x height 318 x depth 268 mm
– Exterior dimensions: width 585 x height 784 x depth 514 mm
– Setting temperature range: 20 – 300ºC
– Working temperature range: at least 10 above ambient temperature to +300°C
– Setting accuracy temperature: up to 99.9°C: 0.1 / from 100°C: 0.5
– Control unit: SingleDISPLAY.
– Ventilation: forced air circulation
– Interior: Stainless steel
– Maximum number of inserts: 6
– Maximum loading of chamber: 50 kg
– Maximum load per tray: 10 kg
– Factory calibration certificate: +160°C, air flow according to ISO 188
– Door: fully insulated stainless steel door with 2-point locking
– Voltage: 230 V, 50/60 Hz, 2000 W
Included:
+ Works calibration certificate: incl. works calibration certificate for 1600C (Memmert)
+ Stainless steel grid, 1 pc
+ Manual